Specifications of system 1. Maximum extracting pressure: 500 Bar. (7,250PSI) 2. Maximum extracting temperature: 80℃. 3. Extraction vessel(Basket is included) : 5L x1 Safety Protection 1. Rupture disk / Safety relief valve. 2. Over-pressure auto shutdown. 3. Jacket insulation protect with the heating part. Control and monitor system 1. Main Control Valve: High Pressure Manual type needle valve. 2. Flow rate monitor: Display on flow meter. 3. Liquide level monitor: Reflex liquid level gauge. 4. Temperature & Pressure Monitor and control system: color Touch Panel HMI and pressure Gauges. Main unit: 1. Extraction Vessel: x1 -MAWP= 500 Bar. (7250 PSI) -MAWT=80℃ -Volume=5L 2. Separation Vessel-1: x1 -MAWP= 172 Bar. (2500 PSI) -MAWT=80℃ -Volume=2L 3. Separation Vessel-2: x1 -MAWP= 95 Bar. (1200 PSI) -MAWT=80℃ Volume=2L 4. CO2(L) pump: 5. Co-Solvent pump: 6. Cooling System: Cooling capacity 2.4KW 7. Heating system: Electric heating. Piping and structure: 1. Piping: seamless stainless steel tube 2. Material of structure and plate: SUS304. System utility requirement 1. Power Supply: 3-PH/220V/ 30A. 2. Compressed air: 7~8Kgf/cm2 3. Dimensions: (L)210cm x (D)125cm x (H)190cm. |