1.Specifications 1-1 Pressure=Max. 7250 Psi / 500 Bar 1-2 Temperature=Max. 100 ℃ 2-1.Extraction Vessel= 1L X1set 2-1-1 Pressure=Max. 7250 Psi / 500 Bar 2-1-2 Temperature=Max. 100 ℃ 2-1-3 Volume=600mlX1 (inner basket) 2-1-4 Material=Stainless steel. 2-1-5 Heating system=electric heating 2-2.Collection system=1 set 2-2-1 CO2 flow rate control valve X 1set 2-2-2 Control valve heating module X 1set 2-2-3 Design for Collecting in atmospheric 2-3 Pressure control system= PM-10000C X 1set 2-4 Cooling system= 1 set 3.Material of tanks, piping and valves. 3-1 Piping:SUS 304 3-2 Structure: Aluminum 4.Control system and monitor 4-1 Control valve=manual type/ high pressure valve 4-2 Temperature control= Temperature controller with thermal couple 4-3 Pressure control=Pressure gauge/Pressure Transducer 4-4 CO2 flow control:flow meter / display the gas CO2 flow rate 5.Safety and Protection 5-1 Rupture dick= set pressure 8500 Psig) 5-2 Heating area with isolation jacket 6.Utility requirement (Supplied by buyer) 6-1 Power require= 220 V, 12 A 6-2 Compressed air: 7Kgf/cm2; 4 M3/min. 6-3 CO2 supply=Liquid CO2 cylinder with inner tube 7.Option( not standard ) 7-1 Spare parts=(ex: seal/rupture dick,…….) 7-2 Co-Solvent pump module ------------------------------------------------------------------------------------------------------------------ Digital Pressure Control Module (PM-10000C) Applied Fluid: Liquid carbon dioxide (LCO2) Specifications 1. System output pressure: 1000~10000 psi (depending on driving pressure) 2. System output flow rate: 200ml/min (up to 4350 psi) 3. Booster pump 4. Built-in pre-cooler (without chiller) 5. Piping: seamless stainless steel tube 6. Certification: Whole system pass CE, EN 61010-2-010(2nd Ed) for use with EN 61010-1:2001 Standard 7. Dimensions: 360 mm × 600 mm × 270 mm 8. Weight: 30 kg Control System 1. Color Touch-Panel for Pressure Control Systems and human-machine operation interface 2. Programmable pressure control system 3. Proportional precise pressure control 4. Multi-step for booster and immediately adjust / response control 5. Alarm (notice) for abnormal operation and operation timer Safety Protection 1. Overpressure abnormal alarm devices 2. Emergency stop devices and automatic gas cut-off installations for overpressure 3. Safety relief valve for overpressure (full flow rate) Utility Needs 1. Power supply: 110/220V 2. Source air: (5~8 kg/c㎡) 3. Chilly: water temperature≤5℃ |