Applied Fluid: Liquid carbon dioxide (LCO2) Specifications 1. System output pressure: 1500~7700 psi (depending on driving pressure) 2. System output flow rate: 456ml/min (up to 4350 psi) 3. Booster pump 4. Built-in pre-cooler (without chiller) 5. Piping: seamless stainless steel tube 6. Certification: Whole system pass CE, EN 61010-2-010(2nd Ed) for use with EN 61010-1:2001 Standard 7. Dimensions: 360 mm × 728 mm × 270 mm 8. Weight: 40 Kg Control System 1. Color Touch-Panel for Pressure Control Systems and human-machine operation interface 2. Programmable pressure control system 3. Proportional precise pressure control 4. Multi-step for booster and immediately adjust / response control 5. Alarm (notice) for abnormal operation and operation timer Safety Protection 1. Overpressure abnormal alarm devices 2. Emergency stop devices and automatic gas cut-off installations for overpressure 3. Safety relief valve for overpressure (full flow rate) Utility Needs 1. Power supply: 110/220V 2. Source air: (5~10 Kg/c㎡) 3. Chilly: water temperature≤5℃ |